iio:health:afe4404 Fix timestamp alignment and prevent data leak.
[ Upstream commit f88eccca ] One of a class of bugs pointed out by Lars in a recent review. iio_push_to_buffers_with_timestamp assumes the buffer used is aligned to the size of the timestamp (8 bytes). This is not guaranteed in this driver which uses a 40 byte array of smaller elements on the stack. As Lars also noted this anti pattern can involve a leak of data to userspace and that indeed can happen here. We close both issues by moving to a suitable structure in the iio_priv() data with alignment explicitly requested. This data is allocated with kzalloc so no data can leak appart from previous readings. Fixes: 87aec56e ("iio: health: Add driver for the TI AFE4404 heart monitor") Reported-by: Lars-Peter Clausen <lars@metafoo.de> Signed-off-by: Jonathan Cameron <Jonathan.Cameron@huawei.com> Acked-by: Andrew F. Davis <afd@ti.com> Signed-off-by: Jonathan Cameron <Jonathan.Cameron@huawei.com> Signed-off-by: Sasha Levin <sashal@kernel.org>